nav_sec3_new.jpg (651 bytes)
Description   |   Features   |   Size   |   Outputs   |   Specifications   |   Mechanical

SNP-X20T
Medical 200W/250W 4x8 open frame

View Specifications in PDF Format
Medical  200W/250W 4x8 open frame
250W with 10 CFM forced air

 

Description Go to top


These 200W, medical grade power supplies are designed with Skynet’s patented Ring-Free ZVS technology to ensure high efficiency and optimal size. This cool running series has no trouble operating at 70 degrees C ambient and has plenty of power in reserve to run at 250W continuously with only 10 CFM forced air cooling.

SNP-X20 series offers extremely high efficiency and is built with long-life components for enhanced reliability. This series features low leakage current and FCC & VDE class B EMI. Meanwhile, it complies with a wide variety of safety regulations so that it can be used in Medical, Industrial, Information Technology or other equivalent applications.

Features Go to top

• Both Medical & ITE Safety Approvals
• Active PFC
• Efficiency between 85% and 88%
• Operation from 0°C to 70°C by convection
• Only 1.5 inch height
• Remote Sense
• Safety Agency Approvals
UL 60950, UL 60601-1
CSA C22.2 No. 60950, CSA 601-1, CUL
TUV EN60950-1, EN60601-1
• EMI
FCC class "B"
EN55022 "B"
• EMS
EN61000-4-2,-3,-4,-5,-6,-11

Size Go to top

4 x 8 x 1.5 inch
102.3 x 201 x 38 mm

DC Outputs Go to top

Output48V
Rated Load5.2A /7.9A pk
Regulation±1%
Ripple150mV
Remote Sense

Specifications Go to top
View Specifications in PDF Format

Input VAC
Voltage, V 90 to 264
Frequency, Hz 47 to 63
Current, A inrush I <30A at 115VAC; <60A at 230VAC
Power Factor > .93 - Active PFC
Harmonics EN61000-3-2 class D
Cooling 200W Free Air Convection
Power Good N/A
Holdup Time >25 mS typical at rated load and 115VAC
Transient consult factory
Efficiency 85%~88% depending on model
Shorted output Protected auto recovery
Over Load Protection Auto-recovery
Short Circuit Protection Auto-recovery
Over Voltage Protection Latch-off
EMS EN61000-4-2, -3, -4, -5, -6, -11
Temperature -20°C to 70°C; derating: 2.5% / °C > 50°C
EMI EN55022-B
Safety Agency UL 60950-1, UL 60601-1
CSA C22.2 No.60950-1, 601.1
EN 60950-1, 60601-1

Notes
Go to top
1. The total output current is rated load with free air convection and max. load with 10CFM of forced air flow over the unit.
2. At factory, all outputs in 60% rated load condition, each output is checked to be within the accuracy range while the main output is setting to within the specified accuracy range at rated load.
3. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
4. Load regulation is defined by changing ±40% of measured output load from 60% rated load at another output set to 60% rated load.
5. Ripple & noise is measured by using 15MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and nominal line.
6. Hold up time is measured from the end of the last charging pulse to the time which the main output drop down to regulation limit at rated load and nominal line.
7. Efficiency is measured at rated load and nominal line.

 

Mechanical Go to top

 

 

Click here to view a printer friendly mechanical drawing.
Medical  200W/250W 4x8 open frame
Return to product index

The information and specifications contained in this data sheet are believed to be accurate at the time of publication. If errors are found, they will be corrected. Consequently, specifications may be subject to change without notice.

Home  l   Products  l  Support   l  Sales  l  About Us  l  Contact Us  l  Site Map